Description
Technical Specifications
| Specifications | Technical Parameters |
|---|---|
| Testing Principles | Through reflection photometer to evaluate the color change of urinalysis strip, then calculate the concentration of related items and output semi-quantitative results |
| Testing Wave Length | 470nm, 525nm, 625nm |
| Testing Seed | 60 samples/hour |
| Report Method | LED; Serial Output; Internal 57mm Micro Heat-sensitive Printer, Chinese and English |
| Display | 320x240LED |
| Input Method | Touch Screen |
| Report Storage | 10000 Reports |
| External Interface | Standard RS-232, can be connected with PC |
| Working Voltage | AC 220V±22V 50Hz±1Hz DC5.85V±0.45V (AA×4) |
| Power consumption of Input | 35 VA |
| Dimensions | 200mm×137mm×52mm |
| Weight | 600g |
| Repeatability | (CV,%)≤1.0 |
| Accuracy | Less than one grade compared to the reference value |
| Stability Contamination | Variable coefficient(CV,%)≤1.0 The negative samples test results will not show the positive after the test of the highest positive sample |
Working and Storage Condition
| Parameters | Working Condition | Storage Condition |
|---|---|---|
| Temperature | 10℃~30℃ | -20℃~50℃ |
| Relative Humidity | ≤80% | ≤93% |
| Atmospheric Pressure | 76-106Kpa | 76-106Kpa |
| Working Condition | Stable Table, avoid direct strong light and strong electromagnetic interference source | Clean, Ventilative, Stable |